Please use this identifier to cite or link to this item:
http://ir.futminna.edu.ng:8080/jspui/handle/123456789/3186
Title: | Modelling of Serpentine Continuous Flow Polymerase Chain Reaction Microfluidics |
Authors: | Abubakar, Mohammed Adeniyi, Akinola A. Abdulkadir Baba, Hassan |
Keywords: | PCR, CF-PCR, DNA Amplification |
Issue Date: | 3-Mar-2012 |
Publisher: | International Journal of Engineering Science and Technology |
Abstract: | The continuous flow Polymerase Chain Reaction (PCR) microfluidics DNA amplification device is a recent discovery aimed at eliminating the cyclic hold experienced while using the alternative stationary device. The Application of Computational Fluid Dynamics is increasingly growing and can help achieve optimal designs before actual fabrication. This paper presents a CFD modelling of a continuous flow serpentine PCR device with narrow and wider channels. There are two temperature regions at 950 C and 600 C for denaturation and annealing respectively. Extension is achieved along the middle of the channel at 720 C owing to temperature gradient. The model require a pressure of 42.6KPa for a 30 cycle amplification |
URI: | http://repository.futminna.edu.ng:8080/jspui/handle/123456789/3186 |
Appears in Collections: | Mechanical Engineering |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
cvpp11_Modelling_of_SerpentineContinuosPCR_MicroFluidics (1).pdf | 214.15 kB | Adobe PDF | View/Open |